12.1.2 OUT commands
OUT commands set device parameters. Remote control mode must be active.
Parameter settings Parameter
Setting
out_sp_00
xxx.xx
Setting of the setpoint for temperature control
out_sp_03
xxx.xx
Setting of the high temperature
out_sp_04
xxx.xx
Setting of the low temperature
out_sp_06
xxx.xx
Setpoint temperature setting via watchdog function
out_sp_07*
x
Set pump to preset levels 1…4:
1 = 55%, 2 = 65%, 3 = 75%, 4 = 85%
out_sp_10
xxx.xx
Setting of the actuating variable via the serial
interface
out_sp_11
x
Setting of the temperature units:
0 = °C
1 = °F
out_sp_12
x
Setting the pressure unit:
0 = bar
1 = °F
out_sp_13
x
Adjusting the flow unit:
0 = l/min
1 = gpm
out_sp_27
xxx
Pump capacity in %
out_sp_28
xxx.xx
High temperature alarm limit setting
out_sp_29
xxx.xx
Low temperature alarm limit setting
* Do not use for new programming, see out_sp_27. Pump stage 4 has an output pressure of approx.
450 mbar.