Initial operation
53
7.5 Set high temperature safety function
Before each new temperature application, the temperature must be set for the
tank and internal reservoir high temperature cut-off. Set a value that is at least
25 K below the flash point of the bath fluid being used. The surface temperature
of the bath fluid must not exceed the flash point at any time. An alarm is
triggered when the set value is exceeded.
Slotted screwdriver, size 3
The device is connected.
The language is set.
1. Switch the device on.
If no temperature liquid has been filled yet, the low liquid level alarm is
displayed.
Depending on the default setting, the high temperature cut-off alarm is
displayed.
2. Call up the
<Main menu>
.
3. Call up the submenu
<Settings>
then menu option
<Customize home screen>
.
4. For one of the possible display values, activate
[High temperature cut-
off].
5. Call up the Home screen.
6. Swipe sideways to show the Home screen view that displays the high
temperature cut-off value T
Protect.
7. Use the screwdriver to set the high temperature cut-off.
The display shows the set value for T
Protect
.
8. Set a value that is at least 25 K under the flash point of the bath fluid being
used.
The set value is immediately active.
9. Turn the device off so that the alarm message is reset.
The high temperature cut-off is set.