Initial operation
55
7.8 Set chiller mode
For a refrigerated circulator, the chiller mode is preset ex works to automatic
operation. Three different chiller modes can be selected in the operating menu:
• Automatic mode: The controller calculates the duration of the heating
process in the event of a setpoint change of more than +5°K. If the
calculated duration of the heating process is longer than 2 ° minutes, then
the refrigeration unit is switched off. The refrigeration unit is started as soon
as cooling is required.
• Always on: The refrigeration unit is permanently switched on during
operation.
• Always off: The refrigeration unit remains switched off during operation.
The device is switched on.
The cooling machine is connected.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Install unit>
then the menu option
<Chiller mode>
.
The menu option can only be called up if the thermostat detects a
connected refrigeration unit.
3. Select the desired chiller mode.
The chiller mode is set.