background image

- 11 -

Flexible Sheath Resistance and Thermocouple Temperature Sensors FlexiTEMP

®

 60

7. Ordering information
7.1 Ordering table

Type

 Description

  T1060

 Sheath resistance temperature sensor

  T1560

 Sheath thermocouple temperature sensor

Code

 Temperature sensor

 Resistance (RTD)

Sheath material

Max. temperature of use

04

 1xPt100, two-wire inside wiring 

1.4404 

up to 500 °C

06

 1xPt100, four-wire inside wiring 

1.4404 

up to 600 °C

06HT

 1xPt100, four-wire inside wiring 

Inconel 600

up to 700 °C

- only for code F7

07

 2xPt100, three-wire inside wiring 

1.4404 

up to 600 °C

08

 2xPt100, two-wire inside wiring 

1.4404 

up to 500 °C

09

 2xPt100, four-wire inside wiring 

1.4404 

up to 600 °C

…VR

 Increased resistance to vibration and shock

up to 500 °C

- only for code 06 F2 … S5

 Thermocouple (TC)

Sheath material

Measuring range

21

 1x"J" (Fe-CuNi), insulated

1.4541 

-200 to +800 °C

61

 2x"J" (Fe-CuNi), insulated, isolated junctions

1.4541 

-200 to +800 °C

22

 1x"K" (NiCr-NiAl), insulated

Inconel 600

-200 to +1100 °C

62

 2x"K" (NiCr-NiAl), insulated, isolated junctions

Inconel 600

-200 to +1100 °C

23

 1x"N" (NiCr-NiAl), insulated

Inconel 600

-200 to +1100 °C

63

 2x"N" (NiCr-NiAl), insulated, isolated junctions

Inconel 600

-200 to +1100 °C

22HT

 1x"K" (NiCr-NiAl), insulated

Nicrobell/Pyrosil

-200 to +1300 °C

62HT

 2x"K" (NiCr-NiAl), insulated, isolated junctions

Nicrobell/Pyrosil

-200 to +1300 °C

23HT

 1x"N" (NiCrSi-NiSi), insulated

Nicrobell/Pyrosil

-200 to +1300 °C

63HT

 2x"N" (NiCrSi-NiSi), insulated, isolated junctions 

Nicrobell/Pyrosil

-200 to +1300 °C

…U

 Grounded version of junction TC

99

 Other

Code

 Accuracy class

Inside wiring material

Measuring range

 Resistance (RTD) according to EN 60751

F1

 B  

Cu  

-50 to +300 °C

- not for code 06HT

F2

 B  

Cu  

-70 to +500 °C

- not for code 06HT

F3

 B

Ni

 1)

-200 to +600 °C

- only for codes 06, 07 and 09 

F7

 B

Ni

 1)

-200 to +700 °C

- only for code 06HT

F4

 A

Cu  

-30 to +300 °C

- only for codes 06, 07 and 09 

F5

 A

Cu  

-100 to +450 °C

- only for codes 06, 07 and 09 

F9

 Other

 Thermocouple (TC) according to EN 60584-1 

T7

 2

T6

 1

- not for TC “N“ with code KV

T9

 Other

Maximal reference temperature for continuous operation

 Sheath

RTD

TC "J"

TC "K", "N"

Code

 Outer diameter of stem D [mm]

Inconel 600

Nicrobell/Pyrosil

S01

 0.5

- only for single TC

 -

 - 

 -

S11

 0.8

- only for single TC

 - 

 -

 -

S21

 1 

- only for single TC

+260 °C

+700 °C

 -

S31

 1.5

- only for single TC

+440 °C

+920 °C

 -

S41

 2

- only for single TC

+440 °C

+920 °C

 -

S51

 3

+400 °C

+520 °C

+1020 °C

+1100 °C

S61

 4.5 

- only for TC

-

+620 °C

+1100 °C

 - 

S71

 6

+600 °C

+720 °C

+1100 °C

+1200 °C

S99

 Other

Code

 Nominal length of stem L [mm]

L....

 Fill length in mm (min. length 100 mm)

Code

 Cold-end version of stem 

2)

VV

 Flying leads (standard length V=10 mm for stem diameter 1 to 2 mm and V=25 mm for stem diameter 3 to 6 mm)

KS1

 Flat single connector (plug), standard version 

- only for TC with stem diameter 3 to 6 mm

KS2

 Flat double connector (plug), standard version 

- only for TC with stem diameter 3 to 6 mm

KM

 Flat single connector (plug), mini version

- only for TC with stem diameter 1 to 3 mm

KV

 Connected connecting cable (for RTD) or compensating cable (for TC) 

3)

- not for TC 2x"N"

H1

 Aluminium head type MA with ceramic terminal block, housing IP 64 

4)

H1G3/8

 Aluminium head type MA with ceramic terminal block, process connection G3/8“, PN16, IP 64 

4)

- only for stem diameter 3 to 6 mm

H1G1/2

 Aluminium head type MA with ceramic terminal block, process connection G1/2“, PN16, IP 64 

4)

- only for stem diameter 3 to 6 mm

S1

 Flange, diameter 42 mm with ceramic terminal block

- only for stem diameter 6 mm

S2

 Flange, diameter 42 mm with set for mounting of transmitter on flange

- only for stem diameter 3 to 6 mm

S3

 Flange, diameter 42 mm with mounting of selected transmitter

- only for stem diameter 3 to 6 mm

 (necessary specifications of transmitter)

S4

 Flange, diameter 42 mm without terminal block, cable leads

- only for double temperature sensor

S5

 Flange, diameter 42 mm without terminal block, embedded pins (acc. to NAMUR)

K9

 Other

 Connection or compensation cable

                         - compulsory for code KV, optional for codes KS, KM and H1

 

5)

Code

 Length of cable AL [mm]

200

 200

1000

 1000

2500

 2500

5000

 5000

....

 Other   - fill length (step 100 mm)
° ... Marked version can be dispatched up to 5 working days (with calibration up to two weeks)

1)

 … Not allowable to use two-wire connection because of nickel inner wiring.              

2)

 … Ambient temperature at the end of cable sheath (at flying leads outcome, connection of connection or compensation cables, connection of connector

        or sensor head) cannot exceed 100 °C (120 °C short-term).

3)

 … Tolerance of stem length and connection or compensation cables length is equal to the greater value of ±2 % of length or ±20 mm;

        accuracy class for TC wires according to EN 60584-3.

4)

 … Not for double RTD, code 07.

5)

 … In option with code KS or KM, the beginning of compensation wires is with flat connector (female) of specified type, specified connector has to be added

        in ordering code (see optional accessories – code Z2, Z3 or Z4).

TC "K", "N"

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

°

Summary of Contents for FlexiTEMP 60

Page 1: ...600 Nicrobell Pyrosil Sheath diameter from 1 to 6 mm Optional nominal length L 0 1 to 50 m Fast response to temperature changes Flexible stem Optional version of cold junction with flying leads conne...

Page 2: ...ct description 4 3 1 Application 4 3 2 Description 4 3 3 Dimensional drawings 5 4 Installation operation and maintenance 6 4 1 Installation and commissioning 6 4 2 Operation and maintenance 8 5 Produc...

Page 3: ...s Sensors with fiberglass insulation of wires must be stored in a dry environment 1 6 Installation and commissioning During installation commissioning operation and maintenance follow the instructions...

Page 4: ...with the sheath made of stainless steel 1 4404 for resistance sensors 1 4541 for thermocouple J or Inconel 600 2 4816 Nicrobell Pyrosil for thermocouple K and N Resistance sensors are supplied with si...

Page 5: ...zalit mi v vody dle NAMUR S2 S3 S4 p evodn k H1 hlavice p ipojovac rouben 42 M4 33 L D 42 M4 33 L D L 50 33 D L 62 33 D G1 2 G3 8 44 33 30 8 M4 Parametry proveden krytu dle po adavku instalovan ho p...

Page 6: ...nce has not been lowered 4 1 3 Special conditions for safe use of sensors in Intrinsically safe version code EI 1 Temperature sensors in intrinsically safe version code EI can be connected to intrinsi...

Page 7: ...Max temperature of measured medium T1 450 C 440 C T2 300 C 290 C T3 200 C 195 C T4 135 C 130 C T5 100 C 95 C T6 85 C 80 C For process temperature measured medium 440 C Tm 1200 C the maximal surface te...

Page 8: ...cal parameters it is necessary to replace the sensor 5 Product specifications 5 1 Technical specifications Resistance sensors type T1060 Measuring resistor RTD 1xPt100 accuracy class A B according to...

Page 9: ...rmocouple wires Steel sheath Hot junction Ceramic insulation Thermocouple wires Steel sheath Ceramic insulation Hot junction Steel sheath Thermocouple wires Ceramic insulation Hot junction Flat connec...

Page 10: ...nsmitter 44 mm span of fixing screws M4 33 mm Applicable transmitters For range of transmitter see category optional accessories in transmitters for head mounting and the catalogue transmitters For ap...

Page 11: ...00 C 520 C 1020 C 1100 C S61 4 5 only for TC 620 C 1100 C S71 6 600 C 720 C 1100 C 1200 C S99 Other Code Nominal length of stem L mm L Fill length in mm min length 100 mm Code Cold end version of stem...

Page 12: ...or PK1 Lock anti pull up cable for standard connectors for single sensor PK2 Lock anti pull up cable for standard connectors for double sensor PK3 Lock anti pull up cable for mini connectors for singl...

Page 13: ...meter sheath 3 to 6 mm M04 M20x1 5 only for sensors with diameter sheath 3 to 6 mm M05 M12 only for sensors with diameter sheath 3 to 6 mm only for shift pipe union PB G01 G1 8 only for sensors with d...

Page 14: ...14 Flexible Sheath Resistance and Thermocouple Temperature Sensors FlexiTEMP 60 8 Appendix 8 1 Declaration of conformity explosion hazard environment version codes EI...

Page 15: ...JSP s r o Raisova 547 506 01 Ji n Czech Republic 420 493 760 811 export jsp cz www jsp cz Your Supplier...

Reviews: