background image

- 7 -

Flexible Sheath Resistance and Thermocouple Temperature Sensors FlexiTEMP

®

 60

failure is not considered and  is sufficient if the power passed 
into  the  sensor  does  not  exceed  0.05  W.  Information  about 
maximum power passed into the sensor can be found in the 
operating manual of the transmitter or associated apparatus.
8. During installation it should be taken into account, that the 
equipment does not meet requirements EN 60079-11, strength 
test 500 V

.

9.  Maximal  temperature  of  process  (measured)  medium  for 
explosion  hazard  environment  of  gases  and  temperature 
classes are listed in the following table.

Temperature 

class

Maximal 

surface temperature

Max. temperature of 

measured medium

T1

450 °C

440 °C

T2

300 °C

290 °C

T3

200 °C

195 °C

T4

135 °C

130 °C

T5

100 °C

95 °C

T6

85 °C

80 °C

For  process  temperature  (measured  medium)  440  °C  <  Tm 
≤  1200  °C,  the  maximal  surface  temperature  of  sensor  Tx 
is  determined  from  maximal  temperature  of  the  process 
(measured medium) Tm and safety addition 10 °C. 
Tx = Tm + 10°C
10.  Maximal  surface  temperature  Tx  for  dust  explosive 
atmosphere is equal to measured medium temperature Tm.
Tx = Tm 

4.1.4 Electrical connection

RTD cable sensors

RD  - red
WH - white
BK  - black
YE  - yellow

1x Pt100/2-wire

1x Pt100/4-wire

2x Pt100/2-wire

2x Pt100/3-wire

RD

RD RD

RD

RD

RD RD

BK BK

WH

WH

WH

WH

WH

WH

YE

RTD with head MA

RD  - red

TC cable sensors

WH - white
BK  - black
GN  - green
PK  - pink

1xJ

BK

1xK

1xN

2xJ

2xK

2xN

WH

GN

WH

PK

WH

1x Pt100/2-wire

1x Pt100/4-wire

2x Pt100/2-wire

RD

RD

RD

RD

RD

+

BK

+

GN

+

PK

+

BK

BK

+

+

GN

GN

+

+

PK

PK

+

Summary of Contents for FlexiTEMP 60

Page 1: ...600 Nicrobell Pyrosil Sheath diameter from 1 to 6 mm Optional nominal length L 0 1 to 50 m Fast response to temperature changes Flexible stem Optional version of cold junction with flying leads conne...

Page 2: ...ct description 4 3 1 Application 4 3 2 Description 4 3 3 Dimensional drawings 5 4 Installation operation and maintenance 6 4 1 Installation and commissioning 6 4 2 Operation and maintenance 8 5 Produc...

Page 3: ...s Sensors with fiberglass insulation of wires must be stored in a dry environment 1 6 Installation and commissioning During installation commissioning operation and maintenance follow the instructions...

Page 4: ...with the sheath made of stainless steel 1 4404 for resistance sensors 1 4541 for thermocouple J or Inconel 600 2 4816 Nicrobell Pyrosil for thermocouple K and N Resistance sensors are supplied with si...

Page 5: ...zalit mi v vody dle NAMUR S2 S3 S4 p evodn k H1 hlavice p ipojovac rouben 42 M4 33 L D 42 M4 33 L D L 50 33 D L 62 33 D G1 2 G3 8 44 33 30 8 M4 Parametry proveden krytu dle po adavku instalovan ho p...

Page 6: ...nce has not been lowered 4 1 3 Special conditions for safe use of sensors in Intrinsically safe version code EI 1 Temperature sensors in intrinsically safe version code EI can be connected to intrinsi...

Page 7: ...Max temperature of measured medium T1 450 C 440 C T2 300 C 290 C T3 200 C 195 C T4 135 C 130 C T5 100 C 95 C T6 85 C 80 C For process temperature measured medium 440 C Tm 1200 C the maximal surface te...

Page 8: ...cal parameters it is necessary to replace the sensor 5 Product specifications 5 1 Technical specifications Resistance sensors type T1060 Measuring resistor RTD 1xPt100 accuracy class A B according to...

Page 9: ...rmocouple wires Steel sheath Hot junction Ceramic insulation Thermocouple wires Steel sheath Ceramic insulation Hot junction Steel sheath Thermocouple wires Ceramic insulation Hot junction Flat connec...

Page 10: ...nsmitter 44 mm span of fixing screws M4 33 mm Applicable transmitters For range of transmitter see category optional accessories in transmitters for head mounting and the catalogue transmitters For ap...

Page 11: ...00 C 520 C 1020 C 1100 C S61 4 5 only for TC 620 C 1100 C S71 6 600 C 720 C 1100 C 1200 C S99 Other Code Nominal length of stem L mm L Fill length in mm min length 100 mm Code Cold end version of stem...

Page 12: ...or PK1 Lock anti pull up cable for standard connectors for single sensor PK2 Lock anti pull up cable for standard connectors for double sensor PK3 Lock anti pull up cable for mini connectors for singl...

Page 13: ...meter sheath 3 to 6 mm M04 M20x1 5 only for sensors with diameter sheath 3 to 6 mm M05 M12 only for sensors with diameter sheath 3 to 6 mm only for shift pipe union PB G01 G1 8 only for sensors with d...

Page 14: ...14 Flexible Sheath Resistance and Thermocouple Temperature Sensors FlexiTEMP 60 8 Appendix 8 1 Declaration of conformity explosion hazard environment version codes EI...

Page 15: ...JSP s r o Raisova 547 506 01 Ji n Czech Republic 420 493 760 811 export jsp cz www jsp cz Your Supplier...

Reviews: