In the event that you experience a malfunction other than those described above
contact qualified staff (retailer, Manufacturer).
This section summarises the most common problems that may arise when using the unit. Before
contacting customer services, carry out the checks listed below.
8 - oPeRaTIonal cHecK anD TRoUblesHooTInG
PRobleM
PossIble
caUse
solUTIon
The green LED indicating device operation
is switched off.
The unit is not
powered.
Check that the switch is in position “I”.
Check that the plug is connected to the
mains power socket.
Check that the plug is connected to the
power socket on the device.
Check that the power socket to which the
device is connected works.
Check that the lid is fitted properly and
locked into place.
Make sure that the fuses next to the
socket in the relevant door are not blown.
If necessary, contact your local retailer.
No sizzling noise comes from inside the
device and the yellow LED that indicates
that maintenance is required is blinking.
The ionising
generators and fan
mesh filter need
cleaning.
Follow the instructions in section
“7.2 - Routine Maintenance”.
No sizzling noise comes from inside the
device and the yellow LED that indicates
that maintenance is required is on and
steady.
The ionising
generators need
replacing.
Follow the instructions in section
“7.3 - Extraordinary Maintenance”.
No sizzling noise comes from inside the
device and the yellow LED that indicates
that maintenance is required is off.
The ionising
generators are
faulty.
Follow the instructions in section
“7.3 - Extraordinary Maintenance”.
Unplug the device from the mains and contact qualified staff for assistance even in the event of
one or several of the cases below:
• The power cable is damaged or worn.
• The plug is damaged or worn.
• Water or liquid has been poured onto the device.
• In the event that there is a malfunction despite all the installation procedures have been carried
out properly.