PAGE 75 of 185
CX0540 Rev 00.1
holder until the latch is fully exposed and snaps out to lock
(Fig. 175)
.
2.
Loosen the vertical adjustment knob and slide the probe holder down until
the wedge is approximately 6 mm
(¼ in)
above inspection surface.
3.
Tighten the vertical adjustment knob
(Fig. 176)
.
4.
Lift the yoke slightly and press the latch button
(Fig. 177)
, then slowly lower
towards scanning surface to apply spring pressure to the wedge
(Fig. 178)
.
TIP:
If less spring force is desired, refer to step 2 and place the wedge
approximately 20 mm (¾ in) above inspection surface.
5.8.4.3 Probe Holder Transverse Adjustment
To adjust the probe holder’s transverse angle, follow these steps:
1.
Ensure the probe holder is in latched, upper position
(Fig. 175)
.
Fig. 177 - Press latch button
Fig. 178 - Lower toward scan surface
Fig. 179 - Loosen 3 mm screw
Fig. 180 - Rotate and tighten
Fig. 181 - Stop post locates 90°