5.8.4.5 Probe Holder Left/Right Conversion
To reverse the probe holder, follow these steps:
NOTE:
To perform this operation the 1.5 mm hex wrench (Fig. 79) is required.
1.
Ensure the probe holder is in latched, upper position
(Fig. 175)
.
2.
Using the supplied 3 mm hex driver
(Fig. 76)
, unscrew the yoke pivot screw
and remove yoke
(Fig. 185)
.
3.
Loosen the probe holder arm adjustment knob and the arm clamp screw.
Slide the probe holder arms off the yoke
(Fig. 186)
.
4.
Flip the yoke 180° and reverse the probe holder arms
(Fig. 187)
.
5.
Place the pivot buttons on the inside of the probe holder arms
(Fig. 188)
using
a 3/8 in wrench
(Fig. 77)
.
Fig. 185 - Unscrew yoke pivot screw
Fig. 186 - Remove probe holder arms
Fig. 187 - Flip yoke and reverse arms
Fig. 188 - Attach arms & move buttons