32 | BA-40100-02-V17
TOPAX
®
DX
Multichannel controller
this stage keep the sensor in the relevant measuring water or buffer
solution (see chapter “5 Measuring values inputs” on page 18).
Enter the requested reference value for the water / solution: The first
digit of the set value is highlighted in grey and can be adjusted.
Use the arrow keys to set up the value:
◀
/
▶
switch over between the individual positions
▲
/
▼
modify numerical values
Confirm and save with the “OK” key.
Confirm the measuring water extraction with the “OK” key.
When calibrating free chlorine and the pH-value the sensor s pulse rate
is shown at the connection to the calibration.
Special feature for the calibration of the pH value
The actual voltage of the single-rod measuring cell AND the design volt-
age of the pH buffer solution are indicated on the display during calibra-
tion of the pH value. These two values allow the operator to monitor the
accuracy of the measuring electrode.
The deviation between the measurement of the pH value with the pho-
tometer and the electrometric measurement can be compensated with
the "offset compensation" option.
mg/l
mV
pH
039 702 719 09
0
ESC
pH value
enter offset
new value pH 7,07
acknowledge with (OK)
0,05 pH
input 2
mg/l
1.2.2.6
Hilfe
Fig. 29: Main menu >> calibration >> pH value >> offset
19.1.5 Menu 1.3: Trend representations
mg/l
mV
pH
Zoom-
Zoom+
ESC
039
702
719
09
0
mg/l
1.3
Anzeige
24 Std.
Fig. 30: Main menu >> trend
The actual trend of the measurements is represented by coloured
curves. These can be constructed at different time intervals. A time
range of 60 minutes, 4 hours or 24 hours can be displayed. To change
the desired time to a respectively bigger or smaller interval use the
„Zoom +“ and „Zoom -“.
19.1.6 Menu 1.4 Status display
mg/l
mV
pH
next
preset
ESC
039 702 719 09
0
status controller
mg/l
1.4.1
W1: 0,45 mg/l
Xp1: 50 %
W2: 7,10 pH
Xp2: 20 %l
Y1: 25 %
Tn1: 25 min.
Y2: 20 %
Tn2: 25 min.
Fig. 31: Main menu >> status
The status-display shows the most significant data for accurate predic-
tion of TOPAX DX proper functioning.
Menu
Status displays
1.4.1
Main parameters of all configured controllers
1.4.2
Switching status of all outputs
1.4.3
Physical values and calibrated resistivity values of all inputs
1.4.4
• used software version
• serial number
• hardware identification data
• input identification
• outputs identification
• number of operating hours
By pressing the “next” key, the TOPAX DX switches to menu 1.4.1 to
1.4.4.
Save and activate preselects
It is also possible for certain applications to save the adjusted values
and activate them when needed. For this presettings several names can
be allocated:
2 Set 2
mg/l
mV
pH
∞C
ESC
039 702 719 63
2
1 Set 1
define
3 Set 3
aktivate
preselects
4 (Autosetup)
1.4.8
Fig. 32: Main menu >> status >> presettings
Press the pre-adjust key and the device switches to menu 1.4.8. Select
from one of the four presettings that you want to activate or overwrite
with the actual values. To select press the arrow keys
▲
and
▼
.
To activate an existing presetting press the activate key. Confirm the
entry with the “OK” key.
To save the actual entries press the "apply" key. The device goes to
menu 1.4.8.1: