BA-40100-02-V17 | 31
Operating instructions
19.1.3 Menu 1.1: Setpoints
controller
mg/l
mV
pH
039 702 719 09
0
pH value
disinfection
ESC
comb. cl
% Salt
setpoint
mg/l
1.1
Fig. 26: Main menu >> setpoint
With the menu "setpoint" each control unit assigns the desired setpoint
to each output. According to the configuration the setpoints can be fixed
for the following control units:
Menu
Controller
Key
1.1.1
Disinfection controller
Disinfection
1.1.2
Controller pH value
pH value
1.1.3
Controller combined
chlorine
bonded chlorine
1.1.4
Controller conductivity
% salt
Select a control unit to set up the setpoint. The device goes to the rele-
vant menu 1.1.1-4.
The first digit of the setpoint is highlighted in grey and can be adjusted.
Use the arrow keys to set up the value:
◀
/
▶
switch over between the individual positions
▲
/
▼
modify numerical values
„OK"
Press OK to save
19.1.4 Menu 1.2: Calibration
mg/l
mV
pH
039 702 719 09
0
ESC
disinfection
pH value
REDOX
temp.
next
calibration
input
mg/l
1.2
Fig. 27: Main menu >> Calibration
mg/l
mV
pH
039 702 719 09
0
ESC
total chlorine
conductivity
input
calibration
mg/l
1.2
Fig. 28: Main menu >> calibration >> next: Page 2 of callibration (with more
than five inputs)
With the Menu Calibration the sensors are calibrated with the inputs.
According to the setting the following inputs can be selected:
Menu
Input
Key
1.2.1
Disinfection
Disinfection
1.2.2
pH value
pH value
1.2.3.3
Redox
Redox
1.2.4.3
Temperature
Temperature
1.2.7.3
Total chlorine measuring
electrode
Total chlorine
1.2.8.3
Conductivity measuring system Conductivity
Select one of the inputs for calibration. The device goes to the relevant
menu 1.2.1-1.2.8.3
The type of calibration (1-point calibration or 2-point calibration) de-
pends on the attached sensor (see also chapter 5), which was defined
in the start-up configuration.
Input
1-point calibration
2-point calibration
Disinfection
Amperometric electrode
*)
Menu 1.2.1
potentialstatic electrode
Menu 1.2.1.3
Encapsulated electrode
*)
Menu 1.2.1
pH value
pH single-rod measuring
cell
- without specifications
about slope rate
Menu 1.2.2
pH single-rod measuring
cell
- with specifications about
slope rate
Menu 1.2.2
Redox
Menu 1.2.3.3
Temperature
Menu 1.2.4.3
Total chlorine measuring
electrode
Menu 1.2.7.3
conductivity measuring
system
Menu 1.2.8.3
*) For start-up the sensors require a 2-point calibration. For inspection/
post-calibration a 1-point calibration is sufficient. Therefore, menu 1.2.1
offers both types of calibration.
Please note the messages on the display. After the calibration has been
selected the TOPAX DX requests one and/or two reference values. At