A9.4) Abnormal and known hardware-software bugs
WARNING:
The following list abnormal behaviour that will hopefully be fixed one day…
A9.4.1) Wrong EDS energy calibration
Starting in April 2019, we have observed a few times a shift in the X-ray peak position.
This is due to the EDS hardware losing its calibration. This is NOT visible on low X-ray
energy lines (e.g., O K
a
or Si K
a
) and should be observable at higher energy (> 5 keV).
Check the position of the Fe K
a
X-ray line (or any other line > 5 keV) and ensure it
matches the theoretical position (Fe K
a
= 6.40 keV; Fe K
b
= 7.06 keV).
If you observe a significant shift in energy (> 0.1 keV, see
Fig. A9.3.1
), call for assistance,
or perform the
reset of the Thermo NSS box
as described below:
1) Select the acquisition mode “Spectrum” in the left menu.
2) Open the tab “Detector Status” on the bottom right panel.
3) Click on “Advanced Status”. It launches the program “FrontEnd Status”.
4) Close the software “NSS”.
5) Click on the button “Reboot” in “FrontEnd Status”. The connection light turns red
(disconnected). Locate the white box behind the Thermo PC (behind the table); wait
5-10 seconds until the indicator light on the top of this box turns green.
6) In “FrontEnd Status” click on “Reconnect”. The connection light should turn green in
the software. If not, you have clicked too early (the hardware was not fully
restarted). Wait and try again. If you still cannot connect, ask for assistance.
7) Restart the Thermo NSS program.
8) Re-run an analysis and check the sum. Ask for assistance if you still observe a shift
in the energy of a high-energy X-ray line.
Summary of Contents for JSM-6390 LA
Page 2: ......
Page 14: ...10 J M Allaz March 14 2021 Figure 10 Opening the sample chamber to remove or place a sample...
Page 18: ...14 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Page 23: ...SEM manual ETHZ v1 4 19 Figure 13 Complete procedure for beam alignment...
Page 30: ...26 J M Allaz March 14 2021 Figure 18 Creating a new NSS project or opening an existing one...
Page 32: ...28 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Page 38: ...34 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Page 48: ...44 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Page 52: ...48 J M Allaz March 14 2021 Figure 27 Processing and extracting element maps...
Page 53: ...SEM manual ETHZ v1 4 49 Figure 28 Calculating and extracting phase maps...
Page 62: ...58 J M Allaz March 14 2021 A6 Thermo NSS toolbars from the NSS manual...