G.2) Activating the CL detector
Once the CL detector arm is in place, proceed as following to activate the CL detector:
•
Turn ON the IR camera.
•
Move your sample close to the CL detector (usually between 16 and 17 mm for
regular thin sections / round mounts).
Always check the IR camera to avoid a disaster!
•
Turn OFF the IR camera when the sample & detector are in place.
•
Perform a beam alignment using the “SEI” mode (see
Section F.3
).
•
Change the signal to “BEIC” in JEOL SEM.
•
Turn ON the CL detector (red button “POWER” on CL controller box).
IMPORTANT:
Brightness & contrast (B&C) for CL image are first adjusted using the
“Brightness” & “Contrast” buttons on the CL control box!
Once the right B&C is
found on the control box, then adjust the B&C of the JEOL SEM program.
Figure 17b.
Control box for the “Centaurus” CL detector.
You can now start acquiring CL images with the Thermo NSS software:
•
To start Thermo NSS follow the instructions given in
Sections H
.
•
To acquire a single or mosaic image, follow the instructions given in
Sections I
.
WARNING! If the “OVERLOAD” button turns ON (Fig. 17b)
, it means that the detector
is saturated with light. Either you forgot to turn OFF the IR camera or you activated the CL
detector while the SEM is opened. Identify the source of this overload (turn OFF the IR
camera, close and pump the sample chamber), and press the “RESET” button, or turn
OFF the CL detector and ON again. If the problem persists, contact the lab manager.
G.3) Deactivating the CL detector
Any time you perform a
sample change or at the end of your session
, you
MUST
deactivate the CL detector
and
retract it
before removing your sample!
•
Change the signal to “SEI” in JEOL SEM.
•
Turn OFF the CL detector (red button “POWER” on CL controller box).
•
Turn ON the IR camera and
move the stage down to
Z = 30 mm (IMPORTANT!)
.
•
Retract the CL detector (see
Fig. 17
, just do it all backwards
J
).
•
IMPORTANT!
Lock the arm when the CL detector is fully retracted!
If you cannot
insert the lock, it is likely because the arm is not fully retracted.
•
You can now safely do a sample change or remove your sample to end your session.
Summary of Contents for JSM-6390 LA
Page 2: ......
Page 14: ...10 J M Allaz March 14 2021 Figure 10 Opening the sample chamber to remove or place a sample...
Page 18: ...14 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Page 23: ...SEM manual ETHZ v1 4 19 Figure 13 Complete procedure for beam alignment...
Page 30: ...26 J M Allaz March 14 2021 Figure 18 Creating a new NSS project or opening an existing one...
Page 32: ...28 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Page 38: ...34 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Page 48: ...44 J M Allaz March 14 2021 Page left blank intentionally a good place for your notes J...
Page 52: ...48 J M Allaz March 14 2021 Figure 27 Processing and extracting element maps...
Page 53: ...SEM manual ETHZ v1 4 49 Figure 28 Calculating and extracting phase maps...
Page 62: ...58 J M Allaz March 14 2021 A6 Thermo NSS toolbars from the NSS manual...