Instruction Manual T601
JAQUET AG
Revision: 004
47/73
Time constant
Sets a time constant T. The target value is reached after 5xT.
The output current follows an e- Function.
If the time constant is longer than 0 seconds, the measurement time
must be set to 5ms or higher.
7.7.16 Copy Parameter Set
Go to <Configuration>, <Copy from Control A>.
A Parameter set may only be copied from Parameter set A! All
parameter of the submenu <System>, <Math>, <System Limits>,
<Relay / Open collector> and <Analog output> are members of the
parameters sets. Therefore, the following Parameters will be copied:
-
Machine factor 1 and 2
-
Deviation
-
Direction Sensing
-
Analog input range
-
Math function selection
-
Upper and lower limits in System Limits
-
System Limit inversion
-
Inclusion of input in the logical combination
-
Inclusion of sensor alarm in the logical combination
-
Relay / Open Collector assignment
-
Relay / Open Collector latch mode
-
Relay / Open Collector fail safe mode
-
Analog output assignment
-
0..20mA or 4..20mA setting
-
Start and end values
-
Analog output time constant