JADEVER SCALE
SNUG III
14
Ⅶ
. Miscellanies
7-1 Specification of trouble information
Error
informatio
n
Causes Solutions
Err
1. Transportation protecting
screw is not loosed
2. Starting zero exceeds
+/-30% of full capacity.
3. LOAD CELL trouble
1. Loose transportation protecting screw.
2. Check whether there is a too heavy article on
the pan and whether other interference exists,
remove the article form the pan.
3. Perform the linear calibration of weight once
again.
4. Replace LOAD CELL or contact Service Dept.
Err4
EEPROM Checksum error
Re-start the scale; in case this error information
appears once again, weld EEPROM once again
or contact Service Dept.
Err5
The weight of an article
exceeds the maximal
ca9e
Remove the over-heavy article from the pan
promptly, and avoid the similar error from then
to avoid the possible damage.
Err6
The weight for calibration is
wrong.
Please load the corresponding calibration weight
pursuant to the list of calibration weights.
Battery
symbol
Batteries are short of
electricity
Replace it with new one or use AC power supply
(transformer).
7-2 List of unit conversion
1 ct
[MET.CARAT]
= 0.1999694 g
1 lb
[AVORIRDUPOIS POUND]
= 453.59237 g
1 oz
[AVORIRDUPOIS OUNCE]
= 28.349523125 g
1 dr
[AVOIRDUPOIS DRAM]
= 1.7718451 g
1 GN
[GRAIN](U.K)
= 0.06479891 g
1 ozt
[TROY OUNCE]
= 31.1034768 g
1 dwt
[PENNY WEIGHT]
= 1.55517384 g
1 MM
[MOMME] (JPN)
= 3.749996 g
1 tl.j
[HONG KONG JEWELRY TAEL]
= 37.4290018 g
1 tl.T
[TAEL](TWN)
= 37.49995 g
1 tl.H
[HONG KONG TAEL]
= 37.799375 g
1 t
[TOLA] (INDIA)
= 11.6638038 g