b) Attach the other indicator (A) so that the indicator rod comes into contact with the inner end of
the driver coupling half.
This indicator is used to measure angular misalignment.
P
A
Y
X
Figure 14: Dial indicator attachment
2.
Rotate the pump coupling half (X) in order to check that the indicators are in contact with the driver
coupling half (Y) but do not bottom out.
3.
Adjust the indicators if necessary.
4.3.5 Pump-to-driver alignment instructions
4.3.5.1 Perform angular alignment for a vertical correction
Before you start this procedure, make sure that the dial indicators are properly attached for measure-
ment.
1.
Set the angular alignment indicator to zero at the top-center position (12 o’clock) of the driver cou-
pling half (Y).
2.
Rotate the indicator to the bottom-center position (6 o’clock).
3.
Record the indicator reading.
When the
reading val-
ue is...
Then...
Negative
The coupling halves are farther apart at the bottom than at the top. Perform one of
these steps:
•
Add shims in order to raise the feet of the driver at the shaft end.
•
Remove shims in order to lower the feet of the driver at the other end.
Positive
The coupling halves are closer at the bottom than at the top. Perform one of these
steps:
•
Remove shims in order to lower the feet of the driver at the shaft end.
•
Add shims in order to raise the feet of the driver at the other end.
X Y
Shims
Figure 15: Side view of an incorrect vertical alignment
4.3 Pump-to-driver alignment
26
Model 3296 EZMAG Installation, Operation, and Maintenance Manual