Common Commands
Copyright © Itech Electronic Co., Ltd.
76
Chapter15 Common Commands
*CLS
This command clears the error queue and the bits of the following registers:
⚫
Standard Event Register
⚫
Questionable Event Register
⚫
Status byte register
Syntax
*CLS
Arguments
None
Default value
Not applicable
Example
*CLS
Query command
None
Returns
Not applicable
*ESE
This command sets or queries the bits in the Event Status Enable Register (ESER).
The ESER is an eight-bit register that determines which bits in the Standard Event
Status Register (SESR) will set the Event Summary Bit (ESB) in the Status Byte
Register (SBR).
Syntax
*ESE <NRf>
Arguments
0 to 255
Default value
0