Sensor Wiring
Modline 5
77
CyberResearch Inc.
Superverter
TM
Model 285 R-422/RS-485 to RS-232 Converter.
CyMOD
TM
Model CM 4530 USB to RS-232/422/485 Converter with isolation.
(Superverter and CyMOD are trademarks of CyberResearch Inc.)
DGH Corporation
A1000 and 2000 Series (The A1000 will also provide 24 Volt DC power for 3 Sensors)
RS-232/RS-485 Converters and RS-485 Repeaters
Black Box Corporation
RS-422/485 PCMCIA Serial I/O Adapters, Single-Port
Product Code IC114A-R2 (for laptop computers)
See section 6
, page 107, for Modline 5 Sensor communications details
and commands.
4.9.2
TSP Terminal Strip Plates Multi-drop Wiring
These illustrations indicate the proper wiring when using TSP Terminal Strip Plates. A power supply
for each Sensor and TSP pair is shown in the first illustration. This is the preferred method.
A single power supply is shown powering three Sensors in the second illustration. A power supply
should not power more than 2 or 3 Sensors.
Use the shielded cable specified in the previous manual sections and install termination resistors as
required.
Figure 69: Multidrop Sensor Wiring with 24 VDC Supply for each Sensor
To Modline 5 Sensor
To Modline 5 Sensor
Modline 5 MSI Multi-Sensor Interface
or PLC, PC, System
RS-485
Ground
TX-A
TX-B
Termination Resistor
at furthest Box
Power Supply
24 VDC
Power Supply
24 VDC