Sensor Installation
34
Modline 5
Figure 18: Dimensions Sensor with DWD
3.7.3
General Installation Requirements
Sensor should be mounted to provide a minimum of 4 inch (101.6 mm) clearance from the
housing for connecting the cable.
Gently bend the cable into its routing position as shown in the photo below.
Minimum cable bend radius is 3 inches (76.2 mm).
Rotate the Sensor to avoid cable obstructions.
Positioning of any mounting accessory on the Sensor should be as close to the front lens as
possible without extending beyond it. Some accessories require clamping towards the middle of
the body.
Never clamp on the Dirty Window Detector hooded mirror or the DWD section! (Not shown)
Never clamp on the rear focusing section of the Sensor.
Figure 19: Interconnecting Cable
3.8
DWD Installation Requirements
The Sensor with the Dirty Window Detector option requires the use of an ESA – Extension Sleeve
Adapter accessory along with other Modline 5 accessories. The ESA, combined with other accessories,
serves as protection for the DWD optics. The photograph (
) illustrates several of the
accessories.
The ESA is used in the following combinations to provide air purging or to create a dead air space zone
in front of the Sensor optics to help keep them clean. The combinations also provide shielding from
stray radiation emanating from high temperature targets and thermal isolation for the Dirty Window
Detector Components. The APA - Air Purge, used in combinations one and two below, provides the
Dimensions in inches and (millimeters)
Sensor with DWD
Front
Rear