Operation
Modline 5
103
Before turning on the Laser remotely with the DPM Digital Panel Meter Interface or
other devices using RS– 485 Communications, or a remote switch, insure there are no
personnel in the path of the beam.
Display
Instructions
Display and Limits
Manual
Pushbutton
Normal Rear Panel Operation. See next page for Remote Switch
Operation:
Under safe operating conditions, Momentarily depress LASER
pushbutton to energize the Class 2 Laser. Press pushbutton again to
de-energize. Laser will automatically turn off after 20 minutes if not de-
energized.
LED above LASER
Button Lights
Remote Laser Operation (Sensor Firmware Versions 1.08 and Higher)
Remote operation is selected from the COMMS – LASR Menu. The operation for PANL, I IN and SWIT
modes of operation are explained below.
COMM Menu
LASR Mode
Selection
Sensor Rear Panel
LASER Switch
RS-485
LS Command
(ModView, DPM,
MSI)
Peak/Hold Switch
(Remote SPST
Switch Required)
I – IN Current Input
(Remote SPST Switch
Required)
PANL
Operation
Energizes and turns
Laser Off.
Energizes and turns
Laser Off.
Not Enabled for Laser
Operation.
Not Enabled for Laser
Operation.
I IN
Remote Current
Input Operation
Can de-energize Laser
anytime.
Can Energize Laser.
Condition Set One
applies.
Can de-energize Laser
anytime.
Can Energize Laser.
Condition Set One
applies.
Not Enabled for Laser
Operation in I IN
mode.
15 mADC input current
(Switch Closed)
energizes Laser.
Current disconnected
(Switch open) de-
energizes laser.
Condition Set One
applies
SWIT
Remote Pk/Hd
Switch
Operation
Can de-energize Laser
anytime.
Can Energize Laser.
Condition Set Two
applies.
Can de-energize Laser
anytime.
Can Energize Laser.
Condition Set Two
applies.
PkHd Switch operates
Laser.
Condition Set Two
applies.
Not Enabled for Laser
Operation in SWIT mode