Sensor Installation
60
Modline 5
3.12
Other Accessories (Sight Tube, Window, WJ-5 Water Cooling)
These other accessories are described in Product Bulletin PB-0070. Specification and mounting
configuration changes are required for certain items when used with the Modline 5.
Consult the IRCON factory for information about using these accessories with the Modline 5 Sensor.
Other Accessory Model / Part
Compatibility and main interface
AA-3
Requires MFL Flange to interface.
AP-Q, AP-CF
Requires MFL Flange to interface.
Consult Ircon, Inc. with application details.
MF-1, MF-2 Threaded collars for threaded
1-1/2 inch NPT sight tubes and 2-1/2 inch
NPT sight tube.
Requires the MFL Flange to interface. See note below.
PM-2 Pipe Mounts for use with
STSC and STA Sight and Target Tubes
Requires the MFL Flange to interface. See note below.
STM Sight Tubes with Flange Surface
Requires the MFL Flange to interface. See note below.
WA-3 Water Cooling Accessory
Not compatible with Modline 5 Sensors.
Do not use to provide cooling.
WJ-5 Water Cooling Jacket with Air Purge.
A mounting kit, UAA, and MFL is required to install Modline 5 Standard
Sensor in WJ-5. Maximum ambient temperature limited to 160
C (320
F).
Note: Not compatible with Modline 5 Sensor with Dirty Window Detector
Option (DWD) installed.
See installation details below.
Table 12: Compatibility with other Accessories
Focus Sensor with rear rotating focus section. If looking through a sight tube or sight
hole, position, align and rotate the Sensor and Sight tube to center the reticle or laser
image in the field of view.
3.12.1
M5WJ5 – Mounting Kit
The M5WJ5 Mounting Kit is used with UAA and MFL Mounting Accessories to install a Modline 5
Sensor in an Ircon Model WJ-5 Water Cooling Jacket. The jacket provides cooling through a stainless
steel coiled pipe embedded in the jacket walls. A flow rate of 20 gallons per hour (75 liters per hour) of
water with a temperature of 90
F (32
C) or less is required.
An air purge is included on the front of the WJ-5. Clean, dry purge air should be provided at a flow rate
6 ft
3
/min. (0.17m
3
/min.).
The jacket protects the Modline 5 Sensor in ambient temperatures up to 160
C (320F). Complete
specifications, installation and piping instructions are included with the kit. The Modline 5 Sensor cable
is rated to 200
C (400F).
Total Lengths Dimension shown with and without MFL Flange.