Invacare® MK6i™ Electronics
14.2.2
Actuator Control
An actuator control setting MUST be selected to operate powered seating through the driver control.
The actuator control menu determines the method for operation of actuators through a driver control that is accessed through a mode switch or
stand-by select.
This menu appears only when there is a multiple actuator interface box or SANODE installed on the seating system.
ACTUATOR CONTROL
OFF
>
4-SWITCH
LATCH. 4SW
4 SW-2 LEVELS
4SWL-2 LEVELS
MOM.1SW
•
OFF - disables driver control operation of the powered seating
for that drive only.
•
4 SW - Directions of driver command mirror Actuator Selection
choices; Momentary mode.
•
LATCH 4 SW - Same as 4 SW, but in latched mode. First
command initiates actuators, repeated command stops actuator.
•
4 SW - 2 Levels - Recommended for head Controls in RIM mode.
Dedicates left and right commands while turning off forward
(occipital pad) command.
– Mode Switch activates level 1, then level 2
♦
Level 1 - Left driver command = Forward actuator selection,
Right driver command = Reverse actuator selection.
♦
Level 2 - Left drive command = Left actuator selection, Right
driver command = Right actuator selection.
– Standby Select bypasses mode switch requirement and allows
Left command (held down) to cycle through each level.
♦
Level 1 - Forward - Reverse actuator selections
♦
Level 2 - Left - Right actuator selections
•
MOM.1SW - Intended primarily for Head Control drivers
operating with RIM.
– Mode switch cycles through each actuator selection choice.
Right command operates selected choice.
– Standby Select allows left command (held down) to cycle
through each actuator selection choice. Right command
operates selected actuator (Momentary Mode).