Invacare® MK6i™ Electronics
ERROR CODE
SYMPTOM
PROBABLE CAUSE
SOLUTIONS
E203
RIGHT MOTOR
FAULT displays and the
wheelchair does not
drive.
Displayed when a problem with
the right motor is detected.
Right on RWD - Left on CWD
Check motor lock engagement (clutch).
Check motor connection plug.
Verify left/right by switching motor plugs.
If fault follows motor, replace motor.
If fault does not follow motor, replace controller.
E204
REMOTE FAULT
displays and the
wheelchair does not
drive.
This is displayed when the
controller determines an
incorrect configuration.
Replace the controller.
E205
CONTROLLER
FAULT displays and the
chair does not drive.
This is displayed when the
controller fails a power-up
test.
Replace the controller.
E206
CONTROLLER
WRONG REMOTE
displays and the
wheelchair does not
drive.
This is displayed when he
controller has determined an
invalid configuration.
Check all connections and wiring.
E207
CONTROLLER
SETUP FAULT displays
and the wheelchair does
not drive.
This is displayed when the
controller module does
not recognize the MK6
CMPJ+/Display as a valid
device.
Replace the controller.
E208
G-TRAC FAULT is
displayed
The G-Trac module or
controller is not functioning
correctly.
Replace the G-trac module and/or the controller. Please
note that the chair will drive with this error displayed,
however, the G-trac feature is disabled and the chair
performs without the benefit of the G-Trac features.
120
1141471-L-00