Invacare® MK6i™ Electronics
Sip n’ Puff (Pneumatic) Controls
WARNING!
Risk of Injury or Damage
Improper mounting or maintenance of the Sip n’ Puff
control including the mouthpiece and breath tube may
cause injury or damage.
Water inside the Sip n’ Puff interface module may cause
damage to the unit.
Excessive saliva residue in the mouthpiece/straw can
reduce performance.
Blockages, a clogged saliva trap or air leaks in the system
may cause Sip N’ Puff not to function properly.
–
Ensure moving parts of the wheelchair, including the
operation of powered seating, DO NOT pinch or
damage the Sip n’ Puff tubing.
–
Saliva trap MUST be installed to reduce risk of water
or saliva entering the Sip n’ Puff interface module.
–
Occasionally flush the mouthpiece to remove saliva
residue.
–
The mouthpiece/straw MUST be completely dry
before installation.
–
If Sip n’ Puff does not function properly, inspect
system for blockages, clogged saliva trap or air leaks.
As necessary, replace mouthpiece, breath tube and
saliva trap.
Contact your Invacare dealer/provider for more information
about maintaining and troubleshooting the Sip n’ Puff system.
16.2
Diagnostics Menu
>>
JOYSTICK STATUS
FAULT LOG
TILT ACTUATOR
RECLINE ACTUATOR
CM LEG ACTUATOR
VERSION
Joystick Status
The JOYSTICK STATUS menu displays joystick throw settings for
each quadrant when actively moving the joystick inductive.
DRIVE 1 INPUT
FWD
REV
LEFT
RIGHT
0
0
0
0
Fault Log
A professional memory card is required in the MK6i
Programmer to view error code and troubleshooting
information. Fault codes in the Color MPJ+ Joystick are date
and time stamped.
The FAULT LOG displays a history of error codes, including those
intentionally caused during factory testing.
Highlighting the Error code and pressing the INFO key on the MK6i
programmer will display Cause of Error code and troubleshooting
steps to resolve the error.
108
1141471-L-00