CALIBRATION
LCD DISPLAY
DESCRIPTION
HARD PUFF CAL.
MIN
.70 MAX
S
H
HARD PUFF
SOFT PUFF
HARD SIP
SOFT SIP
| |
Calibrates pressures required to activate hard / soft, puff
& sip of commands.
•
Separate screens for each of the 4 pressures.
•
Select command to calibrate. Follow instructions
below. Save on completion.
•
GOAL: Separate “S” & “H” values sufficiently for
easy distinction between Hard & Soft commands –
AND set values low enough to assure they can be
consistently achieved. Values should be between
.10 and 1.28.
•
In a Hard calibration mode, use the up/down arrows
to raise and lower the H value which MUST be met
as the user puffs/sips.
•
In a Soft calibration mode, use the up/down arrow
keys to raise and lower the S value, which MUST be
met as the user puffs/sips.
1141471-L-00
101