74
60126081-C
For more information about driving, refer to
Proportional Driving Mode, page 41
Changing Function Screens
6.2 Navigating User Function Screens, page 31
for
more information about changing the function screens.
For the difference between a function screen and
profile, refer to
4.3 User Function Screen Overview,
. For operating the powered seating functions,
refer to
6.16.1 Through Seating Screens, page 54
.
Operating the Lighting Functions and Horn
Via a utility function screen, you are able to operate the lighting
functions with an external input. The utility function screen is
part of one or more profiles and can be activated like a drive or
seating function screen.
1. Activate the utility function screen.
2. Give input to the front
A
to turn on the horn.
3. Give short input to the right
B
to turn on/off the position
lights.
4. Give short input to the left
C
to turn on/off the hazard
lights.
5. Give long input to the left or right
D
to turn on/off the left
or right turn signal.
Activate a drive function screen to drive normally,
while position lights and hazard lights remain turned
on.
After more than 10 seconds, the turn signals turn off
automatically.
6.19.5 Using the Sip and Puff
WARNING!
Risk of Injury or Damage
Improper mounting or maintenance of the Sip and
Puff control, including the mouthpiece and breath
tube, may cause injury or damage.
Water inside the Sip and Puff interface module may
cause damage to the unit.
Excessive saliva residue in the mouthpiece/straw can
reduce performance.
Blockages, a clogged saliva trap or air leaks
in the system may cause the Sip and Puff
not to function properly.
-Ensure moving parts of the wheelchair, including the
operation of powered seating, DO NOT pinch or
damage the Sip and Puff tubing.
-Saliva trap MUST be installed to reduce risk of water
or saliva entering the Sip and Puff interface module.
-Occasionally flush the mouthpiece to remove
saliva residue.
-The mouthpiece/straw MUST be completely dry
before installation.
-If Sip and Puff does not function properly, inspect
system for blockages, clogged saliva trap or air leaks.
As necessary, replace mouthpiece, breath tube and
saliva trap.
Contact your Invacare provider for more information
about maintaining and troubleshooting the Sip and
Puff system.
For maintenance and cleaning instructions, see
.
Sip and Puff is not the most maneuverable or intuitive
control method and therefore requires a considerable
amount of training. In the early tuning stages. This
training is best done outdoors in an unrestricted but
safe area. Also, the presence of an attendant is
recommended.
Driving
The drive function screens for the Sip and Puff are preset in
latched driving mode. For more information, refer to
LiNX® Control System