SECTION 10—SAFETY INSPECTION/TROUBLESHOOTING
30
Part No 1125031
MK6i Electronics
SPJ™+ w/PSS or SPJ+ w/ACC Joysticks
The joystick information gauge and the service indicator display the type of fault or error
detected by the control module. When a fault is detected, the wheelchair may stop and
not drive. The LEDs on the information gauge may flash in a particular pattern or the
service indicator light will flash. The number or type of flashes indicates the nature of the
error. If multiple errors are found, only the first error encountered by the control module
will be displayed.
Information Gauge Display Diagnostics
SYMPTOM
PROBABLE CAUSE
SOLUTIONS
Wheelchair Power On
but does not drive.
System tilted or reclined
beyond drive lock-out
angle (20°).
Motor lock levers are
disengaged.
Speed set too low.
Battery charger
connected.
Return to neutral position (upright and completely
lowered). Refer to Operating the Powered Tilt System
on page 103.
Contact Invacare if this does not solve the problem.
Engage motor lock levers. See wheelchair base owner’s
manual.
Adjust speed setting.
Disconnect battery charger.
Seating system not
functioning or working
intermittently.
Low batteries.
Faulty electrical
connection.
Seat has been driven
under a heavy load for an
extended period of time.
Charge batteries.
Check all connections. Ensure network connectors are
connected properly and caps are installed on both ends
of the series of network connectors. Refer to
Connecting/Disconnecting Network Connectors on
page 118.
Allow time for the electronics to cool down (Light Duty
Use).
DISPLAY
DESCRIPTION
DEFINITION
COMMENTS
All LEDs are off.
Power is off.
All LEDs are on.
Power is on.
Fewer than three LEDs on
implies reduced battery
charge.
Information
Gauge
Display