7 SETUP/MAINTENANCE
Part No 1143155
109
Invacare® Formula™ CG Seating
4.
Remove the two thumb screws securing the top shroud and the
front shroud/battery retention bracket to the wheelchair base
frame.
5.
Remove the top shroud.
6.
Lifting up and forward, remove the front shroud/battery retention
bracket from the wheelchair base frame.
7.
Disconnect the vent battery connector from the vent battery cable
connector.
8.
Remove the vent battery from the wheelchair base frame.
9.
If necessary, remove the two GP32 batteries from the wheelchair.
Refer to the wheelchair base owner’s manual.
.
FIGURE
19
Removing/Installing the Vent Battery
Vent
Battery
Wheelchair
Base Frame
Front Shroud/Battery
Retention Bracket
Mounting
Screws
Vent Tray Battery Cable
Connector Located Here
Top Shroud