7 SETUP/MAINTENANCE
Invacare® Formula™ CG Seating
114
Part No 1143155
•
3G Storm Series
®
Wheelchairs -
i.
Remove the battery cover from the ventilator battery tray.
ii.
Position the battery onto the ventilator battery tray.
iii. Connect the ventilator battery wiring harness to the battery.
iv. Place the battery cover over the battery.
v.
Secure the ventilator battery strap around the battery cover and ventilator battery tray and clip together.
vi. Connect the ventilator battery harness connector to the ventilator battery harness extension (Detail “A”).
•
TDX SP Wheelchairs Only - Ensure all three 22NF batteries are installed and that the front 22NF battery wiring harness is connected to
the ventilator battery harness connector. Refer to Installing/Removing the Batteries (TDX SP with Vent Tray Option) on page 98.
7.
Wheelchairs with Recline Systems Only - Perform the following steps:
A. Test the recline function of the seating system to ensure that no part of the back assembly interferes with the ventilator tray.
B.
If necessary, contact a qualified technician to adjust the recline down limit to avoid interference with the ventilator tray.
FIGURE
22
Preparing the Ventilator Tray for Use
Strap
Rear Portion
of Buckle
DETAIL “A” - 3G STORM SERIES
WHEELCHAIR VENTILATOR
BATTERY TRAY
Battery Cover
Ventilator
Battery Tray
Ventilator Tray
Ventilator Battery
Strap
Ventilator Tray
Strap