7 SETUP/MAINTENANCE
Invacare® Formula™ CG Seating
90
Part No 1143155
Adjusting Headrest Depth/Direction
1.
Loosen, but DO NOT remove, the three socket screws that secure the offset fixture to the slide tube.
2.
If necessary, reposition the headrest to the desired depth by sliding the headrest towards the front of the wheelchair or towards the rear of the
wheelchair.
3.
If necessary, reposition the headrest to the desired position (headrest will move in any direction).
4.
While holding the headrest in the desired position, securely tighten the three socket screws.
5.
If necessary, adjust the height of the headrest. Refer to Adjusting Headrest Height on page 89.
7.13 Replacing Headrest
1.
Remove the three socket screws that secure the headrest to the headrest mounting bracket.
2.
Position the new headrest on the headrest mounting bracket and secure with the existing three socket screws.
3.
If necessary, adjust the height, depth or direction of the headrest. Refer to Adjusting Headrest Height or Adjusting Headrest Depth/Direction
on page 90.
FIGURE
11
Replacing Headrest
For this procedure, refer to FIGURE
10 on page 89.
Headrest Mounting Bracket
Socket
Screw
Headrest
One style of headrest shown for
clarity. Both styles of headrest attach
the same way.