6 ADJUSTMENT
Part No 1143155
71
Invacare® Formula™ CG Seating
6.6
Adjusting the Tension of the Flip Up Footplate
1.
Loosen the mounting screw on the front rigging angle hinge to
decrease the rotation effort.
2.
Tighten the front rigging angle hinge mounting screw to increase the
rotation effort.
3.
Repeat STEPS 1 and 2 for the other footplate.
FIGURE
10
Adjusting the Tension of the Flip Up Footplate
6.7
Adjusting the Angle of the Cantilever Arm
1.
Adjust the cantilever arm angle as necessary to accommodate the
user.
•
To adjust the arm angle down, turn the adjustment screw
clockwise.
•
To adjust the arm angle up, turn the adjustment screw
counterclockwise.
2.
Repeat STEP 1 to adjust the other cantilever arm (as necessary).
FIGURE
11
Adjusting the Angle of the Cantilever Arm
The tension can be adjusted to increase or decrease
the rotation effort of the flip up footplates.
DO NOT remove the footplate mounting screw.
Footplate
Mounting Screw
Center Mount
Front Rigging
Access to the adjustment screw is on the bottom of
the arm tube.
Back Cane
Arm Pad
Adjustment Screw
Arm Tube