6 ADJUSTMENT
Part No 1143155
67
Invacare® Formula™ CG Seating
Angle Adjustment
1.
Loosen, but DO NOT remove, the two flat screws, washer and
locknuts that secure the footplate to the footplate hinge
(FIGURE
4).
2.
Position the footplate to the necessary angle to accommodate the
user (FIGURE
5).
3.
Retighten the two flat screws, washers and locknuts.
FIGURE
5
Angle Adjustment
Perpendicular and/or Inversion/Eversion Adjustment
1.
Insert a flathead screwdriver through the half clamp on the
footplate (FIGURE
4).
2.
Slowly turn nylon adjustment screw in or out until footplate is
perpendicular to the footrest assembly or the desired inversion or
eversion is obtained (FIGURE
6).
FIGURE
6
Perpendicular and/or Inversion/Eversion Adjustment
For this procedure, refer to FIGURE
4 on page 66
and FIGURE
5.
Observe the angle of the footplate for reinstallation.
The settings for positioning the footplates on the
half-clamps may vary for each footplate.
Side View of
Footplate and
Footrest
Support
Footrest Support
Footplate
For this procedure, refer to FIGURE
4 on page 66
and FIGURE
6.
It is not necessary to remove the footplate to
perform this adjustment.
Front View of Footplate
and Footrest Support
Footrest Support
Footplate