background image

ITABAR Flow Sensor Type 

FTMD 

Operating manual

 

 
 

6.3 Insertion Procedure for FTMD 

1. 

Verify that the Flo-Tap insert-retract 
mechanism are in the position as shown in Fig. 
13. 

2.  Verify that the Flo-Tap instrument valves {8} are 

fully closed and that the unit isolating valve {2} 
is fully open (Fig. 13 and 14). 

3. Initiate probe insertion by rotating the drive 

nuts {10} clockwise as viewed from the top, 
using ratchet wrench. The nuts must be 
tightened alternately, about two turns at time to 
prevent binding resulting from unequal loading. 

 

Continue this procedure until probe contacts the 
opposite side of the pipe or end support 
(FTMD-26) 

4. Inspect the packing gland for evidence of 

leakage. If the unit was ordered with high-
temperature gland, additional adjustment may 
be required at this time. 

8
6
10

3

4

 

Fig. 13 retracted Flo-Tap Sensor 

 

5. Connect instrument lines to the instrument 

valves and to the appropriate meter, recorder, 
transmitter or controller. 

6.  Open the Flo-Tap instrument valves {8}. Then 

purge or bleed the connecting lines and readout 
equipment as required. 

 

6.4 Retract Procedure for FTMD 

1.  Fully close the Flo-Tap instrument valves {8}. 

Then, if required, depressurize and disconnect 
the instrument lines. 

2. Loosen slightly packing gland {6+7} before 

proceeding with retraction. 

3.  Retract the Flo-Tap by rotating the drive nuts 

{10} clockwise as viewed from the top, using 
ratched wrench. The nuts must be turned 
alternately, about two turns a time, to prevent 
binding resulting from unequal loading. 

  Continue this procedure until the probe is fully 

retracted as shown in Fig. 14. 

 

8

6

1 0

3

4

2

7

1

 Fig. 14 retracted Flo-Tap Sensor 

Summary of Contents for Analog Node

Page 1: ...Installation of the ITABAR Sensor 6 6 1 Installation of Type FTMD25 6 6 2 Installation of Type FTMD26 7 6 3 Insertion Procedure for FTMD 8 6 4 Retract Procedure for FTMD 8 7 Installation of the Attach...

Page 2: ...ept for drilling and welding equipment These instructions call for the use of a Mueller DH 5 drilling machine or equivalent which is rated for 80 bar at 35 C with a maximum temp erature of 120 C This...

Page 3: ...1 Determination of Pipe Arrangement For design reasons the pipe arrangement at the installation location has to be known before the sensor is manufactured For horizontal pipe arrangements the instrume...

Page 4: ...flow measurements in a horizontal pipe the ITABAR Sensor must be installed in the lower half of the pipe circumference the connections to the instruments have to be located below the pipe axis Fig 5...

Page 5: ...eous installation location As a general rule regulating valves throttle valves and gate type valves should be installed behind the sensor Note If the recommended straight pipe run lengths are not avai...

Page 6: ...l flanged nipple 8 Drill through the pipe wall in accordance with the instructions supplied with the drilling machine For type FTMD 25 35mm 9 Withdraw the drill past the Flo Tap unit isolating valve C...

Page 7: ...one end around the installed weld socket 4 Wrap the other end of the cord around the pipe so that it forms a loop around the pipe Mark the half way point of the pipe circumference on the pipe 5 Now dr...

Page 8: ...h temperature gland additional adjustment may be required at this time 8 6 10 3 4 Fig 13 retracted Flo Tap Sensor 5 Connect instrument lines to the instrument valves and to the appropriate meter recor...

Page 9: ...followed by a zero point error We recommend you to install the differential pressure lines close together connect hi and lo line heat conducting to maintain equal temperature Attention The differenti...

Page 10: ...ng signal The ITABAR sensors are head measuring devices and will not accurately measure two phase flow ITABAR Flow Sensors for low pressure steam have condensate pots in which the condensate water lin...

Reviews: