SERVICE PLAN
51
E
ng
lis
h
ommended� Ask your medical supplier for that� Please clean additional parts by
means of mild soapsuds and a soft sponge usually used in a household� The basic
corpus of the ORTHOLIGHT seat system is provided with a high-quality resisting
plastic coating� Clean the latter with soapsuds and a soft sponge� Avoid aggres-
sive or rubbing agents, as they may damage the surface of the plastic coating�
Disinfection
Please use water-based products for the disinfection such as, for example,
Terralin, Quartamon Med
or
Sagrotan original concentrate
� Always observe
the manufacturer’s instructions for use�
Service Intervals
ORTHOLIGHT needs to be checked for functionality and defects by a specialist
dealer or us regularly or according to usage, but at least once a year� The nec-
cessary checks are listed in the service plan in chapter
17
of this user manual�
16. Service plan
Important:
Inspections must take place every
6
months!
Pos. Area
Check
Adjustment /
functionality
ok
Damage /
deformation
Exchange /
replacement
1
Function check
Back adjustment
Fixing unit
Arm rest
Head support
Abduction wedge
Therapy table
2
Screw
connections
Screws tightened
Rust spots
3
Appearance
(visual inspection)
Paintwork
Flock coating
Cleanliness
Fabric seams
(position / appearance)
Type plate