![INTELLICYT iQue Screener PLUS Hardware Manual Download Page 21](http://html1.mh-extra.com/html/intellicyt/ique-screener-plus/ique-screener-plus_hardware-manual_2073912021.webp)
17
iQue
®
Screener PLUS Hardware Manual: Part Number 11811 Rev B, Getting to Know the iQue Screener PLUS
Overview
IntelliCyt’s iQue Screener PLUS is a platform for screening suspension cells, beads, microbes and mixtures. The
system consists of a robotic plate rail with an orbital shaker, robotic plate sampler with a contact sensing probe,
peristaltic pump, detector, and ForeCyt screening software to design, run, and analyze screening assays.
Requirements
Power
120 VOLT CONFIGURATION
AC Input: 120 VAC, 50-60Hz, less than 3A
Fuses are 5mm x 20 mm Glass-Tube Fuse, Time Delay, 250 VAC, 5 Amp
220 VOLT CONFIGURATION
AC Input: 220 VAC, 50-60HZ, less than 2A
Fuses are 5mm x 20 mm Glass-Tube Fuse, Time Delay, 250 VAC, 5 Amp
The instrument must be plugged into an IntelliCyt approved uninterruptable power
supply (UPS). The UPS must be plugged into a standard, grounded, or earthed
mains electrical outlet, conforming to local codes. Non-grounded or non-earthed
mains outlet adaptors must not be used. The iQue Screener PLUS is provided with
an approved UPS.
Power Cable Between
UPS and Electrical
Outlet
Minimum ratings: 6A, 250 V, 70°C, PVC H03VVF or equivalent, 3 x 0.75 mm2
cable,
Female Connector Type: IEC 60320 C13,
Male Plug: European Type CEE 7/7, or certified equivalent plug for other countries.
Environment
15–30°C, less than 80% relative humidity (Non-Condensing)
Altitude up to 2000 meters
Transient overvoltage Category II
Condensation Pollution Degree 2
Instrument Size
Height x Width x Depth
(includes bottle station,
does not include
computer and UPS):
66 x 130 x 63 cm
97 x 130 x 97 cm with all doors open
Dry Weight
93kg (205 lb)
Getting to Know the