21
Integra
®
Padgett
®
Electric Dermatome Models B and PI
Sterilization
The
Integra Padgett Electric Dermatome Models B and PI
are provided NON-STERILE and
must be sterilized prior to use. The sterilization cycle parameters outlined below have been
validated for the Integra Padgett Electric Dermatome Models B and PI.
Steam Sterilization:
Cycle
Conditions
Exposure
Temperature
Exposure Time
Dry Time
Wrapped Pre-Vacuum
132°C (270°F)
4 minutes
40 minutes
Wrapped Gravity
132°C (270°F)
15 minutes
50 minutes
Wrapped
Pre-Vacuum-UK Cycle
134°C -0/+3°C
(274°F -0/+5°F)
3 minutes
50 minutes
Wrapped Pre-Vacuum
- French Cycle
134°C -0/+3°C
(274° F - 0/+ 5° F)
18 minutes
50 minutes
NOTE:
Ensure sufficient cooling time after the sterilization process to allow the dermatome
temperature to equilibrate to ambient room temperature.
Never immerse instrument, even to cool. This will result in the unit being damaged and
needing repair.
Note:
It is the sole responsibility of the end user to validate alternate sterilization
methods and cycles that do not comply with the validated sterilization methods
and cycles specified within this manual.
ETO Sterilization:
The Integra Padgett Electric Dermatome Models B and PI have been successfully validated for
ethylene oxide sterilization to the parameters specified below:
For Wrapped Configurations subjected to 100% Ethylene Oxide (EO) Sterilization, the
following parameters apply:
Precondition the chamber to the specified exposure temperature and 55% relative
humidity for a minimum of one (1) hour. The 100% Ethylene oxide concentration is
725 mg/L. Note the specified exposure temperatures and times, as well as Aeration times
as listed in Table 2 below:
Summary of Contents for Padgett B
Page 30: ...29 This page is intentionally left blank...
Page 31: ...30 Mod les de dermatome lectrique B et PI Integra Padgett GUIDE DE L UTILISATEUR FR Fran ais...
Page 60: ...59 This page is intentionally left blank...
Page 61: ...60 Integra Padgett Dermatomo elettrico modello B e PI GUIDA DELL UTENTE IT Italiano...
Page 90: ...89 This page is intentionally left blank...
Page 91: ...90 Integra Padgett Elektrisches Dermatom Modelle B und PI BENUTZER HANDBUCH DE Deutsch...
Page 120: ...119 This page is intentionally left blank...
Page 121: ...120 Dermatoma el ctrico Integra Padgett modelos B y PI GU A DEL USUARIO ES Espa ol...