Verifying if the alarm system is functional:
It is possible to verify if the alarm signals are functional with an ECG simulator
capable of generating ventricular fibrillation signals. It is recommended that the
verification be made by a qualified technician, during the preventive maintenance
(see chapter “Inspection and maintenance”).
Information signals
Eazy PRO has information signals, which can be visual, audio or verbal.
VISUAL INFORMATION SIGNAL
DESCRIPTION
Operational Status Indicator
*
Battery in Charge/Charge Complete Indicator
*
Patient’s BPM
Patient’s BPM detected by the device.**
ECG Wave
Patient’s ECG waveform detected by the device.
RCP Interval Chronometer
Reports the elapsed time since the recommendation of the
CPR procedures. In the manual mode, reports the time
elapsed since the device entered the manual mode.**
Illustrative picture of the current step of the AED
procedure. (AED mode only)
Picture that indicates to operator which stage of the
automatic procedure the equipment is performing at the
moment***
Battery Level Indicator
Indicates the approximate device’s battery level.**
Energy Selected (manual mode only)
Indicates the energy selected by the operator.**
* See section “The equipment”, item “Operational status indicator”
** See section “The equipment”, item “Touch screen”
*** See section “Operating in AED mode”
AUDIO INFORMATION SIGNALS (BEEP)
DESCRIPTION
Metronome for CPR compressions
The device beeps at a rate of 100 per minute, helping the
operator during the CPR compressions.
Failure Indicator
If the device is not operational, it beeps according to the
detected failure.*
* See section “The equipment”, item “Operational status indicator”
IMPORTANT: All audio information signals have a sound level of 80dB(A).