Precautions, restrictions and warnings
54
Electromagnetic emissions
Directives and manufacturer declaration - electromagnetic emissions
The Eazy PRO is intended for use in the specific electromagnetic environment below. The customer or user of the defibrillator is
advised to ensure that it is used in such an environment.
Tests
Compliance
Electromagnetic environment - directives
RF Emissions ABNT NBR IEC CISPR11
Group 1
The Eazy PRO only uses RF power for its internal functions.
Nevertheless, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF Emissions ABNT NBR IEC CISPR11
Class B
The Eazy PRO is suited for use in any establishment.
This includes residential establishments and those directly
connected to the public network of distribution of low voltage
electricity which supply domestic use buildings.
Harmonics Emissions IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions IEC
61000-3-3
Complies
NOTE: It is of paramount importance that the true efficacy of the RF shielding and the true attenuation of the RF filter of the shielded
location are checked to ensure that they meet or exceed the minimum values specified.