Operating in AED mode
31
Step 4
2
3
Treatment
recommended. Don’t
touch the patient!
Push the flashing
button to treat the
patient.
If the need for shock is detected, the shock symbol will blink and the device will ask
the user to press the start button again.
Press “START” button again.
The shock will be delivered.
ATTENTION: the user must not touch the patient or conductive surfaces
in contact with him/her during shock delivery, under risk of suffering a
powerful electric discharge.
ATTENTION: disconnect other equipment which do not have defibrillation
protection before defibrillating the patient.
If clinical scans show that defibrillation is not recommended, the Eazy PRO will
announce: “TREATMENT NOT RECOMMENDED”.
Check whether there were patient’s movements during the analysis. If there were,
restart the process. If not, remove pads and start the CPR (cardiopulmonary
resuscitation) procedure. Details on the next section.
Step 5
After the shock, start the CPR procedure.
See more in section 6 – applying CPR.