Operating in manual mode
5
The Eazy PRO enables manual mode operation, as a conventional
defibrillator.
In this situation, the device does not interfere with the treatment, and the user is
responsible for choosing the energy, charging the charge and delivering the shock.
After confirming the mode change, the Eazy PRO ceases to emit sound and visual
orientation, in addition to the automatic safeguards against shocks.
Press button
to start cycle.
Press button
to deliver
shock.
Place pads
on patient.
Shock delivered: Start CPR
(Cardiopulmonary Resuscitation)
Activate
manual mode.
Select
charge
and charge
energy.
ATTENTION: The use of the manual mode is the user’s entire responsibil-
ity. The use by non-qualified professional may cause severe damage and
even the patient’s death.
34