Precautions, restrictions and warnings
58
ECG analysis algorithm
Databases used for the VF/VT recognition algorithm tests:
• MIT-BIH Arrhythmia Database.
• MIT-BIH Atrial Fibrillation Database.
• MIT-BIH Supraventricular Arrhythmia Database.
• European Society of Cardiology Arrhythmia Database.
• Creighton University Arrhythmia Database.
Test report:
•
Recording Methods: t
he files were acquired via internet through the MIT-BIH
database and used via computer.
•
ECG Rhythm Sources:
MIT-BIH, on http://ecg.mit.edu/
•
Rhythm Selection Criteria:
rhythms were chosen according to notes present in
the MIT-BIH database.
•
Criteria and Annotation Methods:
the rhythms were recognized and annotated
in a separate file. Later they were recognized and compared for sensitivity and
specificity calculations.
VF/VT
Nonshockable Rhythms
Shock INDICATED
A
B
NO shock INDICATED
C
D
Sensibilidade=
A
A + C
Especificidade=
D
B + D