User manual |
Specifications
62
Electrical Isolation:
CF Type
Operating Mode:
Continuous operation
Maximum time from rhythm
analysis beginning to discharge
readiness:
20s
Maximum time from rhythm
analysis beginning to full
discharge readiness:
25s
Isolating Method from Mains
Power Supply:
Flexible cable with a mains plug.
Non-frequent use equipment:
Comply with the requirements for non-
frequent use equipment as specified in
NBR IEC 60601-2-4 standard.
Environmental specifications
Temperature:
Operational: 0 to 50ºC (32 to 122 ºF)
Storage: 0 to 70 ºC (32 to 158ºF)
Humidity:
Operational: 10 to 95% RH, without
condensation
Storage: 10 to 95% RH, without
condensation
Altitude:
Recommended to use in altitudes lower
than 2000 meters.