![Instramed EazyPro User Manual Download Page 11](http://html.mh-extra.com/html/instramed/eazypro/eazypro_user-manual_2070101011.webp)
User manual |
Introduction
11
Source: Sociedade de Cardiologia do Estado de São Paulo – SOCESP, Revista
Socesp V.11, no 2.
Use criteria
The Eazy PRO, as well as any other Automated External Defibrillator, must
only be used if the following circumstances, as a whole, are presented:
•
Unconscious victim or unresponsive to verbal or physical stimulus
•
Not breathing
•
No pulse (for professionals)
Other important considerations regarding the use of the Eazy PRO:
• Not recommended for children under one year old.
• Pacemakers may affect the device’s efficiency.
• Medicines in adhesive form must be removed before starting defibrillation.
• Hypothermic patients may not respond well to defibrillation.
• Once the removal of the patient is started, the defibrillation must be interrupted.
Qualified users
Shall be considered qualified users those who have had training in a recognized
institution in the use of automated defibrillators and CPR techniques -
Cardiopulmonary Resuscitation.
About the manual
The function of this guide is to explain how the Eazy PRO Automated Defibrillator
series works, alerting the user to safety risks.
The information contained in this manual belongs to Instramed and cannot be used
fully, or in part, without expressed written consent.
Instramed has the right to make any changes to improve this guide and the product
without prior notice.
This guide is a part of the Eazy PRO and must be kept for further reference.