![Inogen IO-501 User Manual Download Page 47](http://html1.mh-extra.com/html/inogen/io-501/io-501_user-manual_2064284047.webp)
©2022 Inogen, Inc. All rights reserved.
Page 47 of 356 96-12100-01-01-A
English
15.3.1 GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC IMMUNITY:
The Concentrator is intended for use in the electromagnetic environment of home, institution, vehicle, train, air-
plane, boats and other transport modalities. The user of the concentrator should make sure it is used in such
an environment. During the immunity testing specified below the Rove 6 will continue to deliver oxygen within
specification.
Immunity Test
IEC 60601 Test Level
Electromagnetic Environment Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
6Vrms ISM and
amateur frequencies
The Rove 6 Portable Oxygen Concentrator is suitable for the
electromagnetic environment of typical home, institution,
vehicle, train, airplane, boat and other transportation
environments.
Radiated RF
IEC 61000-4-3
10V/m
80 MHz
to 2.7 GHz
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2, 4, 6, 8 and
15 kV air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Electrical fast
transient/burst
EC 61000-4-4
± 2 kV for power
supply lines
Mains power quality should be that of a typical home,
institution, vehicle or other transpiration and mobile
environments.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
Mains power quality should be that of a typical home,
institution, vehicle or other transpiration and mobile
environments.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0% UT for 0.5 cycle at
0 °, 45 °, 90 °, 135 °,
180 °, 225 °, 270 °, and
315 °.
0% UT for 1 cycle
70% UT for 25/30 cycle
0% UT for 200/300 cycle
Mains power quality should be that of a typical home,
institution, vehicle and other transportation and mobile
environments. If the user of the Rove 6 requires continued
operation during power mains interruptions, it is recommended
that the device be powered from an uninterrupted power
supply.
Power frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
30 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical home, institution, vehicle and various
mobile environments. Power frequency magnetic fields from
common appliances in the home are not expected to affect
the device.
NOTE: UT is the a.c. main voltage prior to application of the test level.
Summary of Contents for IO-501
Page 2: ... 2022 Inogen Inc All rights reserved Page 2 of 356 96 12100 01 01 A ...
Page 52: ... 2022 Inogen Inc All rights reserved Page 52 of 356 96 12100 01 01 A ...
Page 102: ... 2022 Inogen Inc Alle rechten voorbehouden Pagina 102 van 356 96 12100 01 01 A ...
Page 152: ... 2022 Inogen Inc Tous droits réservés Page 152 de 356 96 12100 01 01 A ...
Page 202: ... 2022 Inogen Inc Alle Rechte vorbehalten Seite 202 von 356 96 12100 01 01 A ...
Page 252: ... 2022 Inogen Inc Tutti i diritti riservati Pagina 252 di 356 96 12100 01 01 A ...
Page 302: ... 2022 Inogen Inc Wszelkie prawa są zastrzeżone Str 302 z 356 96 12100 01 01 A ...
Page 352: ... 2022 Inogen Inc Id drittijiet kollha riżervati Paġna 352 minn 356 96 12100 01 01 A ...
Page 353: ... 2022 Inogen Inc Id drittijiet kollha riżervati Paġna 353 minn 356 96 12100 01 01 A Maltese ...
Page 354: ... 2022 Inogen Inc Id drittijiet kollha riżervati Paġna 354 minn 356 96 12100 01 01 A ...
Page 355: ... 2022 Inogen Inc Id drittijiet kollha riżervati Paġna 355 minn 356 96 12100 01 01 A Maltese ...