3 | Scope of delivery, transport, storage
INFICON
16 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
WARNING
Risk of injury due to improper transport of a gas cylinder on the mobile
leak detector
► When mounting the bottle holder on the leak detector, follow the sequence of action
steps as described in the separate mounting instructions.
► Mount the base of the cylinder holder flush with the gas cylinder.
► In order not to impair the stability of the leak detector, transport only small to
medium-sized cylinders (maximum 10 l, maximum cylinder size including protective
cap 98 cm, 200 bar)
► Check the supplied tensioning straps for visible damage.
► Always secure the bottle on the bottle holder with both tension straps.
► Close the valve of the gas cylinder before transport.
► Dismantle a connected pressure regulator before transport.
► Protect the valve of the gas cylinder with a valve protection cap during transport.
► Make sure that any connected spray hose is never under tension.
► Observe the manufacturer's safety instructions for handling gas cylinders.
WARNING
Risk of injury due to electric shock
► Only move the device when the mains cable is not plugged in.
► Check the power cord for damage before each use.
WARNING
Risk of injury from falling or tipping device
If the device is improperly transported, it can fall or tip over and cause serious injuries.
► The mobile leak detector is designed for transport on a flat surface. It is not suitable
for being carried or lifted by a crane. Use the handle to push the device.
► For a short, one-sided lifting at smaller obstacles, hold the handle of the device in
pairs. This can be necessary, for example, to be able to cross a cable without
damaging it.
► If you use an industrial truck such as a forklift to transport the leak detector, always
secure it properly against falling.
Summary of Contents for Fab UL6000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 125 152 ...
Page 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 ...
Page 151: ......
Page 152: ......