4 | Description
INFICON
24 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
1 HDMI interface
• For connection to the HDMI interface of a touch monitor, maximum cable
length 3 m. See also "Use external monitor [
91]".
2 / 3 USB 2.0 Interfaces
• For connecting the USB cable of a touch monitor, maximum cable length
3 m.
• For inserting a USB flash drive with update information "Updating the
89]".
• For inserting a USB flash drive for copying measurement data, see
81]".
• For inserting a WiFi USB adapter for a WiFi connection, see "Operate
leak detector via web browser (wireless LAN) [
139]".
4 Ethernet network interface
• For connection to a LAN network, maximum cable length 30 m (shielded)
The remote control RC1000 is optionally available, see also "Accessories and spare
You can also operate the device using a notebooks or tablets, see also "Operate leak
detector via web browser (wireless LAN) [
4.3.2.1 START button
• To start the measurements.
• Can be operated via the touchscreen of the control panel.
• If you press the START key again during measurement, the maximum leakage
rate (Q
max
) that has occurred since you pressed START again will be displayed.
See also "Design of touchscreen [
• If you press the START button on the control panel during measurement, the
maximum leak rate display is updated.
Function
Touchscreen
Control panel
Starting
START button
Display maximum leak rate
Cannot be operated
via touchscreen
Press the START button
again.
4.3.2.2 STOP button
• To stop the measurements.
• Can be operated via the touchscreen of the control panel.