5 | Installation
INFICON
42 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
NOTICE
Operating system can be hacked via USB or Ethernet
The Linux operating system used in the leak detector is not updated automatically and
can therefore contain security vulnerabilities. This vulnerability may be exploited
through the Ethernet and USB interfaces of the leak detector to provide unauthorized
access to the system.
► Ensure that no unauthorized person has access to these interfaces, for example by
using a USB port / Ethernet port lock.
► In order not to jeopardize the security of your company network, never connect the
leak detector directly to the public Internet. This is applies to connections over
WLAN as well as over Ethernet.
► However, if you want to access the web interface of the leak detector remotely, we
recommend an encrypted Virtual Private Network (VPN) connection. However, we
cannot assume any guarantee for the security of VPN connections, which are
provided by third parties.
Prevention of measurement errors due to leaks in the helium source in
the device surroundings
We recommend that you regularly check around the device to a distance of 10 m all
large helium sources for gross leaks. Use a sniffer line to do this.
Observe the information on transporting the device, see also “Scope of delivery,
Procedure
• So that measurement results are not falsified, use in an area that has a constant
ambient temperature if possible.
• For proper functioning of wheels and brakes, remove the adhesive strips from the
wheels.
• In order not to block the exhaust opening on the lower part of the device, place the
feet of the unit on a firm even surface.
• To easily reach the power switch on the back of the device, ensure there is
sufficient free space behind the device.
• Do not expose the device to direct sunlight.