C. Installation Instructions
1. Workplace Requirements
Location: Indoors only. Any outdoor area where the equipment is to be located should meet all the conditions below.
2. For InBody Carrying Case Users
(1) When opening the box, check to make sure all the following items are inside.
(2) Connect the electrode cables to the InBody S10.
Connect electrodes of RA(Right Arm), LA(Left Arm), RL(Right Leg), LL(Left Leg) to each
corresponding part of the InBody S10. InBody S10 offers two types of electrodes: the Touch Type
electrodes and Adhesive Type electrodes.
(1) Operation Condition
Temperature range
10 ~ 40˚C (50 ~ 104˚F)
Relative humidity
30 ~ 75%
Atmospheric pressure range 70 ~ 106kPa
(3) Connect it to an outlet with a grounding terminal. Equipment may be damaged or malfunction due to
electrical shock, which may result in inaccurate test results.
(4) If InBody S10 experiences electrical interference, the test results may be inaccurate. Do not install near
fluorescent lamps, large AC motor equipment (running machines, refrigerators, air conditioners, compressors),
high-frequency thermal therapy devices, and electric heaters.
(5) Do not connect multiple peripherals to the same outlet. If it is plugged into a power outlet such as an electrical
appliance, disconnect it and plug it into a different power outlet.
(2) Adapter
Manufacturer
BridgePower Corp.
Mean Well Co, Ltd
Model
BPM040S12F07
GSM 40A12-P1IR
Power Input
AC 100 ~ 240V, 50/60Hz, 1.2A(12A-0.6A)
AC 100-240V, 50/60Hz, 1.0-0.5A
Power Output
DC 12V, 3.4A
DC 12V, 3.34A