2. Adhesive Type
Attach the EKG electrodes to the examinee’s hands and feet as shown below.
Then, connect the adhesive electrodes to the EKG electrodes.
(1) Hand electrodes
Find the ulna head(outer wrist bone) by feeling the wrist joint. Imagine or draw a line bisecting
the ulna head, perpendicular to the arm. Place the electrode on this line(on the back of the hand),
with the tab of the electrode pointing away from the body. The other electrode should be wrapped
around the middle finger, with the tab facing away from the body.
Ulna head(outer wrist bone)
Black(V)
RED(I)
(2) Foot electrodes
The black electrode should be attached to the medial malleolus (inside ankle bone) in the same way the
black electrode is attached to the ulna head on the hand. The red electrode should be placed at the base
of the second toe.
Malleolus (inside ankle-bone)
Black(V)
RED(I)
•
Once used, the EKG electrodes should be discarded. Reuse may cause contamination through indirect
physical contact.
•
As EKG electrodes will be in direct contact with the body of the examinee, only use products with
CE certification. We officially recommend 2330 Red Dot Resting Electrode-Tab Style of 3M and RT34
Tab-Electrodes for resting ECG of SKINTACT for the InBody S10.