C. Measurement Posture
Maintaining the right posture during the test is essential to achieve reliable results and accuracy. It’s important
for an examinee to maintain their posture for 10-15 minutes so that impedance can be measured accurately.
1. Lying Posture
(1) It is recommended that the examinee lie down for about 10~15 minutes before the test, so that
body water may be dispersed evenly inside the body.
(2) Make sure the arms do not touch the torso. Spread them naturally to a 15 degree angle away from trunk.
(3) Make sure thighs do not touch each other, and spread legs shoulder-width apart.
Attention
• It is recommended that the examinee stays in the testing posture for about 10~15 minutes.
• Make sure the arms do not touch the torso. Spread them naturally to a 15 degree angle away from trunk.
• Make sure thighs do not touch each other, and spread legs shoulder-width apart.
• Precaution must be taken to make sure electrode cables or tester’s body parts are not making any contact
with steel frames.
• If there is a heating mat (electric blanket) on the floor or mat, be sure to turn off the power and unplug the
power plug when possible.
※
Since InBody S10 is shielded between cable and cable cover, it is not affected by interference between
electrode cables and contact with steel structure of electrode cable. However, in order to obtain more
stable measurement value, take precaution to prevent any contact with steel structure.
Make sure arms do not touch the torso.
Make sure thighs do not touch each other.
S10_manual_ENG_G_220317.indd 27
S10_manual_ENG_G_220317.indd 27
2022. 3. 23. 오후 4:57