Do not place the InBody S10 in a location making it difficult to disconnect the power cord.
•
Do not plug in or pull out the power cord with wet hands. There is a risk of an electric shock.
•
Always use an outlet with a voltage range of 100 - 240. Using outlets outside this range may result in fire
or malfunction.
•
When using a power surge protector, make sure that the outlet or the extension cable has adequate power capacity.
•
Do not disassemble or modify the equipment, including internal parts, without written consent from the
manufacturer. This may cause electric shock or injury, product malfunction, inaccurate results, and will
void the manufacturer’s warranty.
•
If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued safe
use of equipment.
•
Do not directly contact the InBody S10 with any other electronic device when the InBody S10 is on. This may
result in an electric shock.
Warning
DANGER
NOTE
CAUTION
참 조
주 의
참 고
주 의
•
If you have any problems installing your InBody S10, please contact InBody for assistance.
•
If the InBody S10 is not plugged into a grounded outlet, damage due to electric surges or product malfunctions
may occur. This may affect the test results.
•
The test results may be inaccurate if the InBody S10 is under electrical interference. Do not install the InBody S10
near products that generate electrical interference such as fluorescent lights, large AC motor equipment(treadmill,
vibration plate, refrigerator, air-conditioner, compressor, etc.), high-frequency thermal therapy equipment, or
heating appliances. Do not share the power source of the InBody S10 with other electrical devices. This may
affect the test results.
•
When connecting the InBody S10 with other test equipment, turn on the other equipment first. When turning off
other equipment, turn off the InBody S10 first. This is necessary to minimize electrical surges to the InBody S10.
•
Always use the specified adapter provided by InBody, as it is a part of the InBody S10. Using other adapters may
result in a malfunction of the InBody S10.
•
Operation of the InBody S10 2,000m above sea level may affect the weight measurement.
•
Adapter must be arranged so the power may easily be cut off when a problem occurs in the InBody S10.